Tandem Mobi Insulin Pump Recall, FDA Reports 1,248 Serious Injuries, Check Which Serial Numbers Need Replacement

The U.S. Food and Drug Administration (FDA) has classified a recall of Tandem Mobi insulin pumps (Tandem Diabetes Care, Inc.) as Class I, its most serious recall category. The FDA published its public safety notice on November 25, 2026, although Tandem sent its first customer notices on September 23, 2026, so many patients and caregivers are only now seeing the warning.

The defect is false-positive occlusion alarms. The pump can report a blockage when none exists and suspend insulin delivery, which can lead to dangerously high blood sugar. Tandem reported 1,248 serious injuries and no deaths as of October 28, 2026. FDA records list 80,938 devices in commerce, including 77,559 in the United States and 3,379 outside the U.S.

This is not a blanket order for every user to stop using their pump. Pumps with certain serial numbers must be returned and replaced. Other Tandem Mobi pumps may stay in use under updated safety instructions.

Do not stop insulin therapy because of this recall. Contact your diabetes care team to set up a reliable backup insulin delivery plan. If your serial number is on the replacement list, contact Tandem at 877-801-6901.

Tandem Mobi Insulin Pump Recall 2026: Quick Facts

Recall detailInformation
ManufacturerTandem Diabetes Care, Inc. (San Diego, California)
Affected productTandem Mobi Insulin Pumps with Interoperable Technology
ClassificationClass I (highest FDA recall risk category)
Customer notices issuedSeptember 23, 2026
FDA database postingOctober 30, 2026
Latest FDA public noticeNovember 25, 2026
FDA recall number / event IDZ-0316-2027 / 99851
Devices in commerce80,938 worldwide; 77,559 U.S.; 3,379 outside the U.S.
Reported serious injuries1,248 as of October 28, 2026
Reported deathsZero as of October 28, 2026
DefectFalse-positive occlusion alarms
Health riskInterrupted insulin delivery and hyperglycemia
RemedyReplacement for designated serial numbers; updated instructions for all other pumps
Customer support877-801-6901 / [email protected]
Recall statusOpen, classified
Legal settlementNone verified as of November 26, 2026

Sources: FDA recall notice, November 25, 2026 and FDA Class I recall database, Z-0316-2027.

Which Tandem Mobi Insulin Pumps Are Being Recalled?

The FDA notice covers Tandem Mobi Insulin Pumps with Interoperable Technology, but the required action depends on each pump’s serial number. The FDA enforcement record lists these model numbers:

Model numberModel number
10040041014811
10107501014813
10126761014936
10127191016911
10136551016912
10136561016913
10135011016914
10137001018075
10140811018077
10148011018078

A matching model number does not mean your pump must be returned. Replacement eligibility is decided by serial number only.

Tandem Mobi Recall Serial Number Lookup: How to Check Your Pump

The FDA published a spreadsheet of pumps subject to replacement: Download the official FDA replacement serial-number list. If the file does not open, go to the FDA recall announcement and select “List of Affected Product Subject to Replacement.”

  1. Find your serial number. Check the pump’s identification label, the pump information screen in the Tandem Mobi app, your original documentation, or the September 23, 2026 Tandem letter. Do not confuse it with the model number or software version.
  2. Search the spreadsheet. Open the FDA file and search for your complete serial number (Ctrl + F on Windows, Command + F on Mac).
  3. Follow the result.
ResultRequired action
Serial number is on the replacement listArrange a pump exchange with Tandem immediately.
Serial number is not on the listKeep using the pump and follow the updated safety instructions.
Serial number cannot be found or verifiedContact Tandem technical support for confirmation.

An unlisted serial number does not mean your pump is unaffected. The updated use instructions apply to every Tandem Mobi pump.

Tandem Mobi Insulin Pump Recall, FDA Reports 1,248 Serious Injuries, Check Which Serial Numbers Need Replacement

Why Is Tandem Recalling Its Mobi Insulin Pumps?

An occlusion occurs when insulin cannot flow through the delivery system, and the pump normally alarms to warn of a blockage. Tandem’s investigation found excessive friction within the pump’s drive module. Under certain conditions this reduces motor performance enough to trigger an occlusion alarm even though insulin flow is not physically blocked. The FDA lists the recall cause as process change control.

When an occlusion alarm triggers, insulin delivery stops and stays suspended until the user responds and restarts delivery. Repeated or unaddressed interruptions can raise blood glucose and, in severe cases, lead to hospitalization.

This does not mean every occlusion alarm is false. Treat every alarm seriously and follow your user guide’s troubleshooting steps.

What Does a Class I FDA Recall Mean?

Under 21 CFR § 7.3(m)(1), a Class I recall involves a reasonable probability that using the product will cause serious adverse health consequences or death.

ClassificationGeneral risk level
Class ISerious health consequences or death
Class IITemporary or medically reversible harm, or a remote probability of serious harm
Class IIIAdverse health consequences are unlikely

Class I does not automatically mean every device must be discarded or that every user will be injured. Recall remedies can include removal, replacement, repair, software correction or revised instructions. Source: FDA, What Is a Medical Device Recall?

What Should You Do if Your Tandem Mobi Pump Requires Replacement?

  • Request a replacement using Tandem’s official replacement request form, or call 877-801-6901, or email [email protected].
  • Ask Tandem about availability, shipping, return procedures and any costs. The FDA notice does not state a universal reimbursement amount or delivery timeframe.
  • Keep insulin delivery uninterrupted. While you wait, follow Tandem’s interim precautions, check glucose regularly and confirm your backup plan with your healthcare provider.
  • Do not change doses, transfer settings or stop treatment without clinical guidance.

Can You Keep Using a Pump That Is Not on the Replacement List?

Yes. The FDA notice allows users with unlisted serial numbers to continue treatment while following the updated precautions below. Contact Tandem if alarms become unusually frequent, persist after troubleshooting or repeatedly interrupt treatment. Customers who received the correction notice can use Tandem’s recall acknowledgement form.

Safety Steps for All Tandem Mobi Users

  1. Respond to every occlusion alarm. Alarms use tones or vibration and escalate until acknowledged. Acknowledge the alert, inspect the cartridge, tubing and infusion site for damage or blockage, fix any problem and resume delivery. Do not repeatedly dismiss alarms.
  2. Turn on smartphone notifications in the Tandem Mobi app and your phone’s settings, and confirm the app can run in the background. See Tandem’s notification guide. Notifications supplement pump alarms but do not fix the mechanical problem.
  3. Monitor blood glucose regularly for unexpected highs or lows. A cleared alarm does not prove enough insulin was delivered.
  4. Keep backup insulin delivery available as prescribed by your care team and described in your User Guide.
  5. Report problems to Tandem and to FDA MedWatch (see below).

Tandem Mobi Recall Injuries: What Are the Health Risks?

The FDA reports 1,248 serious injuries associated with the malfunction as of October 28, 2026, and no deaths by that date. These figures are manufacturer-reported and are not court findings that the device caused each injury.

Insufficient insulin can cause hyperglycemia and, if severe, diabetic ketoacidosis (DKA), a potentially life-threatening emergency. According to the CDC, warning signs include excessive thirst, frequent urination, nausea, abdominal pain, vomiting, deep or rapid breathing and fruity-smelling breath. Seek emergency care if you have trouble breathing, cannot keep fluids down, have several DKA symptoms or have persistently very high glucose readings.

If you are having a medical emergency, call 911 or go to emergency care. Do not wait for a replacement pump or a reply from customer support.

Source: CDC, Diabetic Ketoacidosis

Is There a Separate Tandem Mobi Android App Recall in 2026?

Yes. FDA records show a separate Class II recall of the Tandem Mobi Android app. It is a different defect from the occlusion alarms.

DetailAndroid app recall
Recall initiated / FDA postingOctober 13, 2026 / November 12, 2026
ClassificationClass II
Recall number / event IDZ-0437-2027 / 100091
Model / affected software1015472 / versions 3.5.4, 3.5.2.2 and 3.5.2.1
Quantity4,086 users
ProblemIncorrect delivery calculation involving the Extended Bolus feature

Under specific circumstances the software may omit the immediate portion of an extended dose without a distinct warning. The correction tells affected Android users not to use Extended Bolus. Other functions, including Standard Bolus, may continue under Tandem’s instructions. Check with Tandem before resuming the feature. Source: FDA Class II record

Some patients may need to check their devices against both recalls.

Does the Recall Include Tandem t:slim X2 Pumps?

The November 25 Class I notice covers Tandem Mobi pumps. It does not establish that other Tandem products, including the t:slim X2, share the same drive-module defect. For background on other device litigation, read Medtronic MiniMed Insulin Pump Lawsuit, Dorn v. Medtronic.

Can You Sue Tandem Diabetes Care Over a Recalled Mobi Pump?

Potentially. Patients who suffered documented harm from a defective insulin pump may have claims under state product liability law. As of November 26, 2026, no nationwide class action settlement, certified class or public compensation program for this occlusion-alarm recall has been verified. An FDA recall is a regulatory safety action, not a lawsuit or a judgment requiring Tandem to compensate every patient.

Possible claimWhat an injured patient may need to show
Manufacturing defectThe device departed from its intended specifications and caused harm.
Design defectThe design was legally defective under applicable state law.
Failure to warnInadequate warnings contributed to a foreseeable injury.
NegligenceA duty was breached, resulting in harm.
Breach of warrantyThe product failed to meet an applicable warranty.

Whether these claims succeed depends on the device, regulatory history, state law and evidence linking the malfunction to the injury. See also Dexcom G7 Lawsuit and FDA Recalls.

Which Federal Laws Apply?

Law or regulationRelevance
21 CFR § 7.3Defines FDA recall classifications
21 CFR Part 803Medical device adverse-event reporting
21 CFR Part 806Reporting of qualifying device corrections and removals
21 CFR Part 820Quality system requirements for device manufacturers
21 U.S.C. § 360hFDA notification and repair, replacement or refund authority
21 U.S.C. § 360kFederal preemption of certain state device requirements
21 U.S.C. § 337Enforcement of the Federal Food, Drug, and Cosmetic Act

Manufacturers must report certain corrections, removals and serious adverse events to the FDA. Identifying a regulation does not establish that Tandem violated it. Federal law may also preempt some state-law claims, and private citizens generally cannot sue solely to enforce the FDCA, so any claim needs an individual evaluation by a qualified attorney. Source: FDA, Recalls, Corrections and Removals

How Much Compensation Could Injured Patients Receive?

There is no established compensation schedule or confirmed settlement amount. Depending on the law and evidence, successful personal injury claims may recover medical expenses, hospitalization, lost earnings, pain and suffering and other legally recoverable losses. A replacement pump is not injury compensation, and accepting one does not ordinarily waive injury claims unless you sign an enforceable agreement doing so. Read any release or settlement document carefully before signing.

What Evidence Should Injured Users Preserve?

  • Pump model and serial number, and recall letters
  • Pump alarm and delivery logs, and CGM records
  • Photos or screenshots of the malfunction
  • Hospital and emergency records
  • Communications with Tandem, and replacement and shipping paperwork
  • Medical expenses and lost income

Do not delay a medically necessary replacement to preserve evidence. If the device must be returned, document its condition and identifiers first and ask an attorney about evidence preservation. Similar issues arise in other recalls, such as the Medline Hospital Bed Recall 2026.

How to Report a Pump Malfunction to the FDA

Patients and clinicians can report malfunctions through FDA MedWatch. Include the malfunction, device identifiers, medical consequences and treatment received. A MedWatch report helps regulators but is not a compensation claim and does not start legal proceedings.

Tandem Mobi Recall Timeline

DateDevelopment
September 23, 2026Tandem issues customer replacement and correction notices.
October 13, 2026Separate Android app correction initiated.
October 28, 2026Reported serious injuries reach 1,248.
October 30, 2026Class I recall record posted in the FDA database.
November 12, 2026Separate Android Class II recall record posted.
November 25, 2026FDA publishes its public safety notice on Mobi pumps.
November 26, 2026Replacement and corrective actions remain ongoing.

Frequently Asked Questions

Are all Tandem Mobi pumps being replaced? No. Specific serial numbers require replacement. All other Mobi pumps follow updated instructions.

How many pumps are affected? 80,938 worldwide, including 77,559 in the U.S. That is the total recall quantity, not the number needing replacement.

How many people were injured? Any deaths? 1,248 serious injuries and no deaths were reported as of October 28, 2026.

Should I stop using my pump immediately? No. Do not stop insulin therapy. Check your serial number and arrange a safe transition or backup plan with your healthcare provider.

Will a software update fix the alarms? The Class I recall concerns a drive-module problem. The announced remedies are replacement and updated instructions, so do not assume a software update will resolve it.

Is the recall related to Dexcom monitors? The Class I recall concerns the pump’s drive mechanism, not a confirmed defect in connected glucose monitors.

Is the replacement free? Tandem has a replacement request process for designated serial numbers. Confirm exchange terms and any costs directly with Tandem.

Is there a class action? None specific to this recall has been verified. Injured patients may have individual product liability options.

Is there a replacement deadline? The FDA notice says to arrange replacement immediately and gives no fixed final deadline.

Official Sources

  1. FDA: Tandem Mobi Insulin Pump Recall
  2. FDA: Class I Recall Z-0316-2027
  3. FDA: Replacement Serial-Number Spreadsheet
  4. Tandem: Pump Replacement Request
  5. FDA: Android App Recall, Class II
  6. CDC: Diabetic Ketoacidosis
  7. FDA: What Is a Medical Device Recall?
  8. FDA: MedWatch

Bottom line: Verify your serial number now, follow the instructions that apply to you, keep reliable insulin delivery in place and seek emergency care if you have DKA symptoms. Patients with documented injuries may wish to get an independent legal evaluation.

Disclaimer: This article provides general information, not medical or legal advice. Follow official recall instructions and consult your diabetes care team before changing insulin treatment. Reported adverse events do not by themselves establish legal liability. Information reflects FDA and Tandem sources as of November 26, 2026 and may change; check the FDA notice for updates.

About the Author

Israr Ahmad is a legal content researcher with 4+ years of experience covering class action settlements and consumer rights cases. He has researched and published coverage of 2,500+ settlements using verified court records, settlement administrator filings, and government sources. Learn more about Israr.